Boost Your Productivity!Translate documents (Ms-Word, Ms-Excel, ...) faster and better thanks to artificial intelligence!
https://pro.wordscope.com
https://blog. wordscope .com
Analytical X-ray equipment
Baggage X-ray equipment
Baggage inspection X-ray equipment
Diagnostic X-ray equipment
Diagnostic X-ray installation
Diagnostic X-ray valve
Diagnostic radiology
Diagnostic x-ray system chair
Radiant therapy
Radiotherapy
Roentgenography
Roentgenology
X-ray analysis equipment
X-ray baggage inspection equipment
X-ray department
X-ray equipment

Traduction de «diagnostic x-ray equipment » (Anglais → Français) :

TERMINOLOGIE
voir aussi les traductions en contexte ci-dessous
diagnostic X-ray equipment

équipement à rayonnement X de diagnostic | équipement de diagnostic à rayons X


X-ray equipment in medical diagnosis part A: recommended safety procedures for installations and use - safety code 20A [ X-rays in medical, dental and paramedical diagnostic radiology - recommended installation and safety procedures ]

Les appareils radiographiques en diagnostic médical, partie A : techniques de sécurité recommandées pour l'installation et l'utilisation - Code de sécurité 20A [ Les rayons X en radiologie diagnostique médicale, dentaire et paramédicale - Recommandations concernant les installations et les techniques sûres ]


analytical X-ray equipment [ X-ray analysis equipment | X-ray equipment ]

appareils d'analyse aux rayons X [ appareils d'analyse à rayons X ]


baggage inspection X-ray equipment [ X-ray baggage inspection equipment | baggage X-ray equipment ]

appareils à rayons X pour l'inspection des bagages [ appareils d'inspection des bagages aux rayons X | appareils à rayons X pour le contrôle radioscopique des bagages ]






diagnostic X-ray installation

installation à radiodiagnostic


diagnostic radiology | radiant therapy | radiotherapy | roentgenography | X-ray department | roentgenology

radiodiagnostic | radiologie diagnostique | radiologie thérapeutique | rayons X | roentgenthérapie | roentgenologie


Automatic-aperture-control diagnostic x-ray system collimator

collimateur de système de radiodiagnostic à commande d’ouverture automatique


Diagnostic x-ray system chair

chaise de système de radiodiagnostic
TRADUCTIONS EN CONTEXTE
12. Diagnostic X-ray equipment, being X-ray devices that are used for the examination of the human body, not including dental X-ray equipment with an extra-oral source that is subject to Part II of these Regulations, photofluorographic X-ray equipment, radiation therapy simulators and computer-assisted tomographic equipment.

12. Appareil de radiodiagnostic : dispositif à rayons X pour l’examen du corps humain, à l’exclusion des appareils de radiographie dentaire à rayonnement X de source extra-orale assujettis à la partie II du présent règlement, de l’équipement de radiographie photofluorographique, des simulateurs de radiothérapie et des appareils tomographiques assistés par ordinateur.


30 (1) If high voltage can appear across the X-ray tube of the diagnostic X-ray equipment, then the radiation emitting from the X-ray source assembly of the equipment must not exceed an air kerma rate of 20.0 µGy/h or an exposure rate of 2.3 mR/h when

30 (1) L’ensemble radiogène de tout appareil de radiodiagnostic ne doit pas, si une haute tension peut apparaître dans le tube radiogène, émettre un rayonnement dont le débit de kerma dans l’air est supérieur à 20,0 µGy/h ou dont le débit d’exposition est supérieur à 2,3 mR/h lorsque :


29 (1) The leakage radiation from the X-ray source assembly of diagnostic X-ray equipment must not exceed an air kerma rate of 1.0 mGy/h or an exposure rate of 115 mR/h when the equipment is operated at the nominal X-ray tube conditions of loading that correspond to the maximum specified energy input in one hour.

29 (1) Le rayonnement de fuite de l’ensemble radiogène de tout appareil de radiodiagnostic ne peut excéder un débit de kerma dans l’air de 1,0 mGy/h ou un débit d’exposition de 115 mR/h, lorsque l’appareil fonctionne à une tension nominale avec application d’une charge correspondant à l’apport énergétique maximal déterminé qui est produit pendant une heure.


8 (1) In the case of diagnostic X-ray equipment, other than mammography equipment, when an object set out in column 1 of an item of the table to this subsection is positioned between the patient and the X-ray image receptor, the aluminum equivalent of the object shall not exceed the amount set out in column 2 of that item, as determined using an X-ray beam that

8 (1) Dans le cas d’un appareil de radiodiagnostic, à l’exclusion d’un appareil à mammographie, tout objet qui est visé à la colonne 1 du tableau du présent paragraphe et placé entre le patient et le récepteur d’image radiologique doit avoir un équivalent en aluminium n’excédant pas la limite prévue à la colonne 2, déterminée au moyen d’un faisceau de rayons X :


For more results, go to https://pro.wordscope.com to translate your documents with Wordscope Pro!
(3) The automatic exposure control device of diagnostic X-ray equipment, other than mammography equipment, when the X-ray tube voltage and the thickness of the objects described in subsection (2) are held constant or varied as specified in columns 1 and 2 of an item of the table to this subsection, must limit the variation in optical density of the resulting radiograms to the quantity set out in column 3 of that item.

(3) Le dispositif de commande automatique d’exposition de tout appareil de radiodiagnostic, à l’exclusion d’un appareil à mammographie, doit, selon que la tension radiogène ou l’épaisseur des objets visés au paragraphe (2) est variable ou constante, tel qu’il est indiqué aux colonnes 1 ou 2 du tableau du présent paragraphe, limiter la densité optique du radiogramme produit à la valeur prévue à la colonne 3.


In particular, as regards the use of X-ray equipment, security scanners like any other equipment using ionising radiation, need to respect the legal provisions set for their use.

Plus spécifiquement, en ce qui concerne les équipements à rayons X, les scanners de sécurité doivent, comme tout autre équipement utilisant des rayonnements ionisants, respecter les dispositions juridiques prévues pour leur utilisation.


1. Manufacturers shall provide unrestricted and standardised access to on-board diagnostic (OBD) information, diagnostic and other equipment, tools including any relevant software and vehicle repair and maintenance information to independent operators.

1. Les constructeurs fournissent aux opérateurs indépendants un accès illimité et standardisé aux informations du système de diagnostic embarqué (OBD), aux équipements de diagnostic et autres, aux outils, y compris les logiciels appropriés, et aux informations concernant les réparations et l'entretien des véhicules.


Can the Commission confirm that the use by the British immigration service of gamma-ray equipment to scan goods vehicles at ports of entry to the UK in order to locate possible illegal immigrants in no way contravenes Euratom regulations designed to protect the public from radiation?

La Commission pourrait-elle confirmer qu'en utilisant des équipements à rayons gamma pour inspecter les camions aux ports d'entrée du Royaume-Uni afin de localiser d'éventuels immigrants clandestins, les services de l'immigration britanniques n'enfreignent en aucun cas les règlements Euratom visant à protéger le public des rayonnements?


2. Points out that controls of liquids in hand luggage by X-ray equipment cannot detect explosives in liquids, and calls upon the Commission and the Member States to intensify their efforts to support research studies to find effective tools for the detection of explosives in liquids;

2. souligne que le contrôle de la présence de liquides dans les bagages à main à l'aide de dispositifs à rayons X ne permet pas de détecter la présence d'explosifs dans les liquides et invite la Commission et les États membres à encourager davantage les recherches visant à la mise au point de bons outils de détection de la présence d'explosifs dans les liquides;


The original text could unintentionally cover other devices (e.g. recording of diagnostic X-ray images on digital media) for which the classification into Class IIa seems to be disproportionate.